Loading...

H80M

Overview

H-80M Series High Flow Oxygen Cannula can be used for patients with spontaneous breathing, including artificial airway respiratory insufficiency, and can be used for patients in hospitals and long-term care institutions. High Flow Oxygen Cannula can be used for patients with spontaneous breathing, including artificial airway respiratory insufficiency, and can be used for patients in hospitals and long-term care institutions. Provide a maximum of 80L / min air-oxygen mixed gas, oxygen concentration can be set automatically, providing an exclusive AutoFlow function, SmartFlow function. Disinfection-free design, Water chamber has a backflow prevention valve to prevent gas backflow.

Description

The HFNC oxygen therapy is carried out using air oxygen blender ,active humidifier single heated tube and nasal cannula it is respiratory support. The HFNC continuous to increase easily set up it may also create positive pharyngeal pressure to reduce work of breathing the device midway classical oxygen delivery system like high concentrations CPAP generators . It is designed a heated humifided mixture of air and oxygen at a high flow than the inspiratory flow. It adjust the flow rate and body weight provides a degree of distending pressure. It also decrease works of breathing and oxygenation . It utilizes a soft large bore nasal prongs it is similar to used on mechanical ventilator with the HFNC. It overcome the leading causes of discomfort seen with conventional oxygen therapy. It is a difficult quality level of PEEP generated by the high flow. It support breathing in people with acute heart failure , Acute respiratory distress syndrome. It avoids rebreathing of co2 gas and therefore decreased anatomize dead space. It promoted a slow and deep breathing . it airway hydrated and it improved mucus clearance . This improves FRC and reduce work of breathing it flows used are high humidification is necessary to avoid drying of respiratory secretions and for maintaining nasal cilia function . It is mild positive airway pressure and lung volume recruitment. It will titrate FIO2 and flow to promote comfort and oxygen saturation. It ensure the monitor spo2 saturation and pleth wave form. It ensure maintain a o2 saturation on NRB mask It delivery of optimally humidified and warm gases It decreased body energy expenditure. It is used for intubation and extubation support as well as treatment and prevention of post operative respiratory failure and for oxygenation patients during trails or bronchoscopy.

Benefits & Specifications

• Advanced dual fan design, two levels of HFlow and LFlow
• Provide a maximum of 80L/min air/oxygen mixed gases, oxygen concentration can be set automatically
• 9-stage humidification temperature: 29℃-37℃, accurate to 1℃
• Two levels of HFlow and LFlow, 2L/min~80L/min, can meet patients of most ages
• Comprehensive monitoring indexes: real-time monitoring of oxygen concentration, flow, temperature, respiratory rate, sPEEP.
• Innovative High Flow model AutoFlow TM Automatically set the output flow parameters based on the peak expiratory flow(PEF) of the patient, simplifying the setting steps, satisfying the treatment effect and improving patient comfort
• SmartFlow Follow the patient's every breath, switch the flow between breaths, provide a higher flow when inhaling, and a lower flow when exhaling, follow your every breath like a bi-level ventilator, comfortable and more economical oxygen resources.
• Hot standby function designed for convenient use and saving oxygen

Specifications

-

Device Size

Dimensions

 

335 mm × 227 mm × 158 mm

Weight:

 

about 4 kg

Water capacity:

To maximum fill line 235 mL

Product Use, Transport and Storage

 

Operation

Transport and Storage

Temperature:

18°C to 28°C (64.4˚F to 82.4˚F)

-25°C to +70°C (-13˚F to +158˚F)

Humidity:

≤ 93% Non-condensing

≤ 93% Non-condensing

Atmospheric Pressure:

760 to 1060 hPa

760 to 1060 hPa

Heated Humidifier

Humidifier Output:

No less than 33 mg/L, at 37 °C target

Maximum Delivered Gas Temperature:

≤ 43°C

Mode of Operation

Continuous

Work Mode

HFlow, LFlow, AutoFlow, SmartFlow

SD Card

SD card can record data and fault information

AC Power Consumption

220 V - 240 V ~, 50 Hz / 60 Hz, 2.5 A

Electric Shock

Type of Protection Against Electric Shock

Class II Equipment

Degree of Protection Against Electric Shock

Type BF Applied Part

Degree of Protection Against Ingress of Water

IP22

Sound Power Level

<30 dB, When the working mode of the device is HFlow, and the output flow is 25 L/min

Maximum oxygen input flow

80 L/min

FiO2

Range: 21% ~ 100 %, binning interval 1%

Margin of Error: ± (5% + 5% set value)

FiO2 Display Accuracy

± (5% + 5% set value)

Temp

Range: 25°C ~ 45°C

Temp Display Accuracy

± 2°C

Flow

Range: 2 L/min ~ 80 L/min

Margin of Error: ± (2 L + 10% set value)

Flow Display Accuracy

± (2 L + 10% set value)

sPEEP

Range: 0 ~ 20 hPa

Margin of Error: ± (1 hPa + 20% set value)

Humidity Compensation

Range: -3~3,-3、-2、-1、0、1、2、3, the higher the range, the higher the humidification output

Flow Step Interval

Range: 15 L/min ~ 80 L/min

Tube

Length: 1.8 m

Margin of Error: ± 10%

The connection should not fall off under the 45 N tension.

Specification of Patient Connector

φ 18 mm cone joint

Heater Plate AC Power Consumption

220 V - 240 V ~, 50 Hz / 60 Hz, 1 A

Video

Interested to